GENERAL INFORMATION

  • These medical devices are intended exclusively for dental or surgical use by qualified healthcare professionals.
  • All instruments are supplied non-sterile and must be cleaned, disinfected, and sterilized before their first use and before each subsequent reuse.
  • Prior to use, all devices should be inspected for damage, contamination, or deformation caused during transport or handling.
  • Protective covers, if provided, must be removed before sterilization.
  • Instruments are designed for specific clinical applications only. Misuse may result in harm to both the operator and the patient.
  • Reprocessing must be performed by trained personnel assigned to maintain and monitor device integrity and safety.
  • Damaged, bent, or corroded instruments must not be used and should be properly disposed of or returned to the manufacturer.
  • Reprocessing life is determined by wear, frequency of use, and reprocessing method; there is no fixed reprocessing cycle limit.
  • Reprocessing equipment (washers, autoclaves) must be maintained as per manufacturer instructions.
  • Do not attempt self-repair. All repairs must be done by certified Swadeshi service personnel.

WARNINGS & SAFETY PRECAUTIONS

  • Always handle sharp and pointed instruments with extreme care.
  • Do not sterilize mixed metals together (e.g., aluminum with stainless steel).
  • Do not exceed 134°C in sterilization.
  • Never use strong alkaline (pH > 9) or acidic (pH < 4) detergents.
  • Avoid cold sterilization unless validated for specific instruments.
  • Do not process titanium, anodized aluminum, mirrors, or coated tips in ultrasonic cleaners unless specifically recommended.
  • Always process jointed instruments in open position.
  • Follow the exact concentration and exposure time provided by the detergent/disinfectant manufacturer.
  • Use medical-grade cleaning and disinfecting agents only.

REPROCESSING CYCLE

Step 1: Collection

  • Collect instruments immediately after use and transport them in a closed, puncture-resistant container to prevent contamination and damage.

Step 2: Decontamination / Pre-Soaking

  • Soak in approved enzymatic or quaternary ammonium-based solutions (e.g., glutaraldehyde-free, neutral pH detergents).
  • Follow solution-specific dilution and soaking time instructions.

Step 3: Rinsing

  • Rinse thoroughly with clean, preferably demineralized water to remove chemical residue.

Step 4: Cleaning

A. Manual Cleaning

  • Use soft, non-metallic (e.g., nylon) brushes.
  • Do not use steel wool or abrasive pads.
  • Clean all surfaces, joints, serrations, and lumens.
  • Ultrasonic cleaning is recommended for hinged and delicate instruments.
  • Place instruments in open position and fully submerged.
  • Avoid overlapping sharp tips and mixing different metals.

B. Automated Washer Disinfectors

  • Use validated programs that include thermo-disinfection.
  • Ensure the equipment is properly loaded and maintained.

Step 5: Drying

  • Dry immediately using lint-free cloths or compressed air.
  • Do not leave instruments wet, as it promotes corrosion.

 INSPECTION & MAINTENANCE

  • After cleaning, inspect visually and functionally:
  • Check for cracks, pitting, rust, or deformation.
  • Test cutting edges, alignment, ratchet function, and joints.
  • Sharpen cutting instruments (e.g., curettes, scissors) as needed.
  • Lubricate all metal-to-metal moving parts using medical-grade lubricant (instrument milk) that is autoclave-safe.
  • Reassemble disassembled instruments before sterilization.

PACKAGING

  • Use medical-grade sterilization pouches or trays.
  • Instruments must be dry and in open (unlocked) position.
  • Protect sharp edges.
  • Label with sterilization and expiry date, and operator initials.

STERILIZATION

  • Recommended method: Steam Autoclave (Class B preferred)
  • Cycle Parameters:
  • Temperature: 134°C (273°F)
  • Exposure Time: 5 minutes
  • Drying Time: 20–45 minutes (varies by device load)
  • Use only demineralized water.
  • Flash sterilization and dry heat sterilization are not recommended unless validated for specific devices.
  • Follow sterilizer maintenance protocols for safe use.

STORAGE

  • Store in cool, dry, and dust-free environments between 5°C–40°C.
  • Keep away from direct sunlight and heat sources.
  • Only store instruments in intact, sealed pouches.

DEVICE-SPECIFIC GUIDELINES

  • Please refer to product-specific inserts (where applicable) for additional handling guidance on items such as mirrors, coated instruments, titanium tips, scalpel blades, and syringe systems.

SYMBOLS USED

  • 🔄 Reusable
  • ♻️ Medical Device
  • 🧼 Washer Disinfector Safe
  • 💨 Steam Sterilizer Safe
  • 🧽 Ultrasonic Cleaning Allowed
  • ⚠️ Handle With Care

CONTACT INFORMATION

For further assistance, reconditioning service, or spare parts:

🏢 Corporate Office & Marketing Office

The Bizz International

Address:
No. 6/3-2F, Campbell Cross Road, Austin Town, Bengaluru – 560047, Karnataka, India

📞 Customer Care: +91-9971599566, +91-9068889566

📧 Email: tbisre2012@gmail.com

🆔 GSTIN: 29AGIPH0143B1ZY

MANUFACTURER & REGISTERED OFFICE

🧾 Manufacturer & Registered Office

Swadeshi Instruments (Acrowndentco)

The Bizz International

Address:
No. 58, Mahmoodpur Tiwai, Dehradun-Delhi Highway, Saharanpur – 247001, Uttar Pradesh, India

📞 Mobile: +91-9971599566

📞 Customer Support: +91-9068889566

📧 Email: tbisre2012@gmail.com

🌐 Website: www.dentalmall.in

🆔 GSTN: 09AGIPH0143B1Z0

🔖 Manufacturing License (MFG): UP/M/MD/018891

🆔 MD42 Number: UP1142MD00001

📌 UAN: UP64A0005321